WSX100/200/300 FAMILY SWITCHMODE POWER SUPPLY SPECIFICATIONS
Specifications:
1. General Character:
The model WSX series power supplies are designed for use in medical, information technology
or laboratory environments. They provide a "universal" AC input operation, and can supply
0 to approximately 100W of stable DC output power. Fan cooling is provided to allow higher power densities than
is possible without fan cooling.
1.1 The model number suffix "M", indicates that a specific model is for use in a medical environment.
2. Electrical Specifications:
2.1 Input: 100-250Vac 50/60Hz 250VA
(90-264 Vac, 47-63 Hz is the tested operational range, 100-250 Vac range is marked on product to comply with safety standards.)
2.1.1 Input Fuses:
A current fuse is incorporated in each of the 2 AC lines, rated at 3.5A, 250V, slo
blo type fuse element, 2AG size, configured with integral welded on leads. Available from Littlefuse, part number
230-3.5.
2.1.2 Input Switch:
An optional input switch may be provided if required. A "I" and a "O"
will be marked on it, or next to it.
2.2 Output: ( Optional suffix letter "M", indicates medical usage. )
Single Output Models
|
WSX150 |
5.0Vdc 16A |
|
WSX116 |
16Vdc 6.3A |
Multi Output Models ( "W" indicates a maximum available combined output power . )
| WSX201 WSX202 WSX203 WSX204 WSX205 WSX206 WSX207 WSX208 WSX209 WSX210 |
+12Vdc 7A +15Vdc 6A +20Vdc 4.5A +24Vdc 4A +5Vdc 12A +5Vdc 12A +5Vdc 12A +5Vdc 12A +5Vdc 12A +12Vdc 7A |
-12Vdc 4A -15Vdc 3.2A -20Vdc 2.4A -24Vdc 2A -5Vdc 4A +12Vdc 5A +15Vdc 4A +20Vdc 3A +24Vdc 2.5A +18Vdc 3.3A |
96W 96W 96W 96W 80W 90W 90W 90W 90W 96W |
| WSX301 | +5Vdc 12A | +12Vdc 3A | -12Vdc 2A | 90W |
| WSX302 | +5Vdc 12A | +15Vdc 2.5A | -15Vdc 1.5A | 90W |
2.3 Output Variance, Voltage:
+/- 5% at Full Load, any input voltage or environmental condition
2.4 Load Regulation, Voltage:
Not to exceed 600 millivolts for 100% to 0% load change, varies due to output cord
selected, remote sense option available, contact factory.
2.5 Protection Circuitry: The output is protected from short circuit and prolonged overload. The internal circuitry is protected against a locked fan condition.
3. Mechanical Specifications:
3.1 Enclosure Dimensions: 6.5"L X 3.75"W X 2.56"H inches, enclosure outline drawings available upon request.
3.2 Input Configuration: Male IEC320, 3 conductor, power inlet.
3.3 Output Configuration: 6 foot length standard, various conductor arrangements and jacket material selections are available. Cordsets may be terminated with various dc connectors, custom output connector assemblies are available, contact customer service for details. Standard in-stock cordsets can be seen in the Jerome Industries engineering manual / catalog on pages 10 and 11.
4. Environmental:
|
4.1 Operating :
|
Please contact customer service if a wider operating range is required. |
4.2 Storage :
Temperature: -40 to 75C
Humidity: 10% to 95% non-condensing
Atmosperic pressure: 50 to 106 kPa
4.3 Liquid Ingress:
Spill-proof per IEC60601-1 clause 44.3, vertical drip proof per IEC60529.
4.4 Ingress Protection Level: IPX1
4.5 Audible noise:
52 dBA maximum white noise measured 12 inches from any side. Virtually no detectable
narrowband sound apparent.
4.6 Fan:
Fan will operate at normal speed from 100% load to approximately (25 to 40)% load.
From no load to approximately (25 to 40)% load fan is stopped.
5. Regulatory Compliance
5.1 Safety Compliance: Regulatory Agency UL UL testing and CSA TUV
These power supplies have been certified by the following safety agencies to the standards
shown:
5.1.1 Single Output Models
These power supplies have been certified by the following safety agencies to the standards shown:
evaluation of
Canadian
requirements.
Standard
UL60601-1
UL544
UL60950-1
C22.2#601-1
C22.2#125
C22.2#60950-1
C22.2#601-1
C22.2#125
C22.2#60950-1
EN60601-1
EN60950-1 Classification
Isolation, BF
Patient Care, Grounded
SELV
Isolation, BF
Risk Class 2G
SELV
Isolation, BF
Risk Class 2G
SELV
Isolation, BF
SELVFile#
E160908
E160908
E148586
E160908
E160908
E148586
LR58275
LR58275
LR58275
5.1.2 Multi Output Models
These power supplies have been certified by the following safety agencies to the standards
shown:
| Regulatory Agency UL |
Standard UL60601-1 UL544 |
Classification Isolation, BF Patient Care, Grounded |
File# E160908 E160908 |
5.2 Power supplies shipped to European market need to bear the CE mark. Jerome Industries
can supply a CE mark under the following standards:
| Standard EN60601-1 EN60950 |
Classification Isolation, BF SELV |
5.3 Electro Magnetic Compatibility:
Meets requirements of European EMC directive and FCC requirements.
5.3.1 Medical:
Meets EMC
requirements from EN60601-1-2 (EMC requirements for ElectroMedical Equipment).
5.3.2 Emissions:
Meets level B requirements of FCC 47 CFR part 15 and CISPR11 (EN55011) and CISPR22 (EN55022),
EN61000-6-3 (light industrial) and line harmonic emissions EN61000-3-2.
5.3.3 Immunity:
Meets CISPR24 (EN55024), EN61326 and EN61000-6-1.
5.3.4 EMC Responsibility:
It is the responsibility of the OEM customer to verify EMC compliance with his system
before marketing the end-use system.
Notes:
1) Emissions are monitored on all prototypes using a spectrum analyzer with peak, quasi peak and average mode power measurement capabilities. Conducted emissions are recorded for both 120 Vac and 230 Vac input conditions. A correlated radiated emissions program is used in order to economically verify the radiated emissions from a specific model.
The emissions program works from the premises of the power supply acting as a dipole antenna radiator, one side of the dipole antenna being the input cord, the other side being the output cord. Levels are monitored on the cords by a common mode current sensing device operating into a spectrum analyzer. Correlation factors have been established which are based upon the difference between radiated field measurements made at independent labs and Jerome Industries lab common mode current measurements on the cords of reference power supplies.
Using this technique, Jerome Industries has experienced a 100% passing rate for prototypes submitted for radiated field emissions testing at independent labs.
OEM System Engineer Instructions
* The EMC characteristics of the final system comprised of both the power supply and the main system must be evaluated to the applicable EMC environment regulations.
* Should a power supply need repair, please return it along with a note describing the particular problem to:
Jerome Industries Inc.
730 Division St.
Elizabeth, NJ 07201
U.S.A.
* Circuit operation block diagrams, isolation diagrams, and a list of safety critical components will be available on request to appropriately qualified personnel.
* All products which contain externally replaceable double input fuses shall contain the following text in the system installation manual: "CAUTION: DOUBLE POLE/NEUTRAL FUSING"
Medical Products
* When used with medical products the Instruction Manual prepared for the
end use product must incorporate at a minimum, the following power supply warning notices:
a) Warning: When cleaning the power unit, use a cloth dampened with cleaning alcohol on the outside of the enclosure only. Do not immerse product in water or a safety hazard could arise during use.
b) Warning: Do not use this power unit in the presence of flammable anesthetics.
* Periodic leakage current testing should be done on the combined power supply and end-use system on a yearly basis when used in a hospital environment where such test equipment is commonly available.
Appendix A
WSX100 Family Leakage Current
Measurements per IEC60601-1 Clause 19
| NC | SFC | |
| Earth Leakage Current | 268 | 503 |
| Enclosure Leakage Current | 1.6 | 2.2 |
| Patient Leakage Current (to output negative cord connection) |
11.52 | 144.1 |
| Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) | N/A | 567.6 |
Notes:
1) Measurement data in microamps
2) Measurements made with 275 Vac, 60 Hz input.
3) Worst case measurement condition, including humidity preconditioning.
4) Reduced leakage designs available (including CF category compliant designs), contact factory for availability.
| Back | Home |