MINI POWER FAMILY SWITCHMODE POWER SUPPLY SPECIFICATIONS
Specifications:
1. General Character:
The model WSC100, WSA400 and WSA500 series power supplies
are designed for use in medical, information technology or laboratory environments.
They provide a "universal" AC input operation.
1.1 These units are available in various international wall plug-in configurations, or with an IEC320 power inlet for easy universal input configurability.
1.2 The model number suffix "M", indicates that a specific model is for use in a medical environment.
1.3 Optional suffix indicates mechanical input configuration:
N, North American (or NEMA) plug-in
U, United Kingdom plug-in BS1363
E, European mainland, CEE7/16
A, Australian, AS3112
If blank, then the standard IEC320 inlet is specified.
2. Electrical Specifications:
2.1 Input: 100-250Vac 50/60Hz
(90-264 Vac, 47-63 Hz is the tested operational range, 100-250 Vac range is marked on product to comply with safety standards.)
2.1.1 Input Fuses:
A current fuse is incorporated in the AC input line,
rated at 1A, 250V, slo blo type fuse element. Fuse(s) are not user accessible. Medical grade
models supplied with a 3 conductor input configuration have a second fuse in the
AC input neutral line as well.
2.2 Output:
WSC133 |
3.3Vdc 1.5A |
|
WSC113 |
13Vdc 0.77A |
|
WSA433 |
3.3Vdc 3.3A |
|
WSA414 |
14Vdc 1.40A
|
|
WSA533 |
3.3Vdc 5.0A |
|
WSA514 |
14Vdc 2.15A
|
2.3 Output Variance, Voltage:
+/- 5% at Full Load, any input voltage or environmental condition, for standard output
cordsets. Tighter tolerance available, contact factory.
2.4 Load Regulation, Voltage:
Fully Regulated Models: Output will stay in +/- 5% window from 100% to no
load. Optocoupler used to feed back control signal to primary
controller.
2.5 Maximum Output Ripple:
3.3 - 6V: 50mVp-p
6.1 - 24V: 100mVp-p
24.1 - 36V: 150mVp-p
36.1 - 48V: 200mVp-p
2.6 Protection Circuitry: The output is protected from short circuit and prolonged overload. The unit is thermally protected from high ambient temperature conditions.
3. Mechanical Specifications:
3.1 Enclosure Dimensions:
WSC100 family: 2.70"L X 2.30"W X 1.56"H inches. WSA400 and WSA500 families:
3.25"L X 2.72"W
X 1.89"H inches, enclosure outline drawings available upon request.
Standard color black, custom colors available, consult factory.
3.2 Input Configuration:
Option 1) Male IEC320, 3 conductor, power inlet. Additional detachable input cordset
must be supplied in addition to power adapter. Standard international configuration, 100-240V rated.
Option 2) Wall Plug-in NEMA 5-15P or 1-15P, Standard North American style for 120V,
also used in Japan and some other area's in the far east. 100-120V rated.
Option 3) Wall Plug-in BS1363 U.K. prong assembly, plastic ground pin. 220-240V rated.
Option 4) Wall Plug-in CEE7/16 standard European mainland 2 conductor. 220-240V rated.
Option 5) Wall Plug-in AS3112 standard Australian 2 or 3 conductor. 220-240V rated.
3.3 Output Configuration: 6 foot length standard, various conductor arrangements and jacket material selections are available. Cordsets may be terminated with various dc connectors, custom output connector assemblies are available, contact customer service for details. Standard in-stock cordsets can be seen in the Jerome Industries engineering manual / catalog on pages 10 and 11.
4. Environmental:
|
4.1 Operating :
|
Please contact customer service if a |
4.2 Storage :
Temperature: -40 to 75C
Humidity: 10% to 95% non-condensing
Atmosperic pressure: 50 to 106 kPa
5. Regulatory Compliance
5.1 Safety Compliance:
These power supplies have been certified by the following safety agencies to the standards
shown:
|
Regulatory Agency UL TUV |
Standard UL60601-1 C22.2#601-1 EN60601-1 |
Classification Isolation, BF Isolation, BF Isolation, BF |
File# E160908 E160908
|
5.3 Electro Magnetic Compatibility:
Meets requirements of European EMC directive and FCC requirements.
5.3.1 Medical:
Meets EMC
requirements from EN60601-1-2 (EMC requirements for ElectroMedical Equipment).
5.3.2 Emissions:
Meets level B requirements of FCC 47 CFR part 15 and CISPR11 (EN55011) and CISPR22 (EN55022),
EN61000-6-3 (light industrial) and line harmonic emissions EN61000-3-2.
5.3.3 Immunity:
Meets CISPR24 (EN55024), EN61326 and EN61000-6-1.
5.3.4 EMC Responsibility:
It is the responsibility of the OEM customer to verify EMC compliance with his system
before marketing the end-use system.
Notes:
1) Emissions are monitored on all prototypes using a spectrum analyzer with peak, quasi peak and average mode power measurement capabilities. Conducted emissions are recorded for both 120 Vac and 230 Vac input conditions. A correlated radiated emissions program is used in order to economically verify the radiated emissions from a specific model.
The radiated emissions program works from the premise that the power supply acts as a dipole antenna radiator, one side of the dipole antenna being the input cord, the other side being the output cord. Levels are monitored on the cords by a common mode current sensing device operating into a spectrum analyzer. Correlation factors have been established which are based upon the difference between radiated field measurements made at an independent lab and Jerome Industries lab common mode current measurements on the cords of reference power supplies.
Using this technique, Jerome Industries has experienced a 99% passing rate for prototypes submitted for radiated field emissions testing at independent labs.
OEM System Engineer Instructions
* The EMC characteristics of the final system comprised of both the power supply and the main system must be evaluated to the applicable EMC environment regulations.
* Should a power supply need repair, please return it along with a note describing the particular problem to:
Jerome Industries Inc.
730 Division St.
Elizabeth, NJ 07201
U.S.A.
* Circuit operation block diagrams, isolation diagrams, and a list of safety critical components will be available on request to appropriately qualified personnel.
* All products which contain externally replaceable double input fuses shall contain the following text in the system installation manual: "CAUTION: DOUBLE POLE/NEUTRAL FUSING"
Medical Products
* When used with medical products the Instruction Manual prepared for the
end use product must incorporate at a minimum, the following power supply warning notices:
a) Warning: When cleaning the power unit, use a cloth dampened with cleaning alcohol on the outside of the enclosure only. Do not immerse product in water or a safety hazard could arise during use.
b) Warning: Do not use this power unit in the presence of flammable anesthetics.
* The Instruction manual must also contain a notice regarding methods of disposal of the equipment at the end it's useful life.
* Periodic leakage current testing should be done on the combined power supply and end-use system on a yearly basis when used in a hospital environment where such test equipment is commonly available.
* This product is classified as Class 2 (double insulated), the optional input functional ground is used only for reduction of leakage current available at the output connections.
Appendix A
WSC100 Family Leakage Current
Measurements per IEC60601-1 Clause 19
| NC | SFC | |
| Earth Leakage Current | 165.1 | 250.4 |
| Enclosure Leakage Current | .84 | 2.26 |
| Patient Leakage Current (to output negative cord connection) |
22.8 | 101.52 |
| Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) | N/A | 269.1 |
| NC | SFC | |
| Enclosure Leakage Current | 1.58 | 2.66 |
| Patient Leakage Current (to output negative cord connection) |
51.5 | 77.5 |
| Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) | N/A | 142.7 |
Appendix B
WSA400/500 Family Leakage Current
Measurements per IEC60601-1 Clause 19
| NC | SFC | |
| Earth Leakage Current | 157 | 231 |
| Enclosure Leakage Current | 1.36 | 2.08 |
| Patient Leakage Current (to output negative cord connection) |
58 | 89 |
| Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) | N/A | 153 |
| NC | SFC | |
| Enclosure Leakage Current | 1.28 | 2.06 |
| Patient Leakage Current (to output negative cord connection) |
64 | 112 |
| Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) | N/A | 204 |
| NC | SFC | |
| Earth Leakage Current | 230 | 339 |
| Enclosure Leakage Current | 1.21 | 1.91 |
| Patient Leakage Current (to output negative cord connection) |
65.7 | 100.5 |
| Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) | N/A | 169.4 |
Notes:
1) Measurement data in microamps
2) Measurements made with 275 Vac, 60 Hz input.
3) Worst case measurement condition, including humidity preconditioning.