75 Watt Medical Grade Automobile Power Adapter

  • Wide Input Voltage Range
  • 75 Watts Maximum Output Power
  • CF Level Isolation for Maximum Patient Safety
  • Car/Truck Input Connector
  • Suitable for Use with All Mobile Medical Devices
    (i.e. Respirator, defibrillator, suction pump, etc...)
 
UL, cUL, TUV Certified CE marked for European Compliance CEC / DOE EISA 2007 Compliant

INPUT SPECIFICATIONS

Input Voltage Range 10.8-17.6VDC Continuous
Extended Input Voltage Range for 60 Seconds 10-26VDC DCL100 & DCL200
10-20VDC DCL300

OUTPUT SPECIFICATIONS

Voltage and Current See Model Charts Page 1
Output Voltage Variance ±5%, includes input & output condition variation and component variation
Ripple & Noise (PARD, Note 1) 1% of Vout to 100mV p-p, max.
Current Limit 110% to 160% of Full Load Non Latching,
Auto Recovery
Over Voltage Protection 110% to 135% of Vout Non Latching,
Auto Recovery Protection

SAFETY & EMC


ENVIRONMENTAL SPECIFICATIONS

Operating Temperature -20°C to 40°C
Derating 20% 40°C to 50°C
70°C for 5 minutes
Storage Temperature Designed to -25 to +75°C
Operating Altitude Designed to 0-4,000M
Storage Altitude Designed to 0-10,000M
Humidity 10% to 95% RH, Non-Condensing

GENERAL SPECIFICATIONS

Isolation 4500VDC
Efficiency 70-90% Output Dependent
Water Ingress (Note 3) Per IEC60529, IPX2

MECHANICAL SPECIFICATIONS

Housing Dimensions 4.75” x 3.00” x 1.90” Excludes Rubber Feet
Input Configuration Standard Auto Plug 1M Length, 18AWG
Output Configuration Various; Custom 3’ to 15’ length cords,
various output connectors available

NOTES
1. Measured with a 0.1μF ceramic and a 10μF electrolytic capacitors across the output terminals. The oscilloscope bandwidth is set at 20MHz a co-axial cable will be used to measure it. The test condition is maximum load.

Regulatory Compliance

Designed for use with medical, information technology, and general purpose equipment to the following standards:

All models are certified to:
UL 60601-1 / 60950-1
CSA 22.2 #601-1 / #60950-1 *
TUV EN 60601-1 / 60950-1
CE LVD/EMC directives,

* cUL Mark

EMC

Emissions: Level B requirements of FCC CFR 47 part 15 and CISPR11 (EN55011) and CISPR22 (EN55022), EN61000-6-3 (light industrial) and line harmonic emissions EN61000-3-2.
Immunity: CISPR24 (EN55024), EN61326 and EN61000-3-2.

Certification records available upon project request.

Model Number Output Voltage Output Current Output Power
       
DCL15.0MC 5 5 25W
DCL16.0MC 6 5 30W
DCL17.0MC 7 4.3 30W
DCL18.0MC 8 3.75 40W
DCL112.0MC 12 3.3 45W
DCL113.5MC 13.5 3.3 45W
DCL115MC 15 3 45W
DCL116MC 16 2.8 45W
DCL117MC 17 2.65 45W
DCL118MC 18 2.5 45W
DCL119MC 19 2.37 45W
DCL120MC 20 2.25 45W
DCL25.0MC 5 7 35W
DCL26.0MC 6 7 42W
DCL27.0MC 7 6.8 48W
DCL28.0MC 8 6 48W
DCL212.0MC 12 4 48W
DCL213.5MC 13.5 4 54W
DCL215MC 15 4 60W
DCL216MC 16 3.75 60W
DCL217MC 17 3.5 60W
DCL218MC 18 3.33 60W
DCL219MC 19 3.16 60W
DCL220MC 20 3 60W
DCL313.5MC 13.5 4.7 63.5W
DCL315MC 15 4.7 70.5W
DCL316MC 16 4.7 75W
DCL317MC 17 4.4 75W
DCL318MC 18 4.17 75W
DCL319MC 19 3.95 75W
DCL320MC 20 3.75 75W