WSZ100/200/300 FAMILY SWITCHMODE POWER SUPPLY SPECIFICATIONS

Specifications:

1. General Character:
The model WSZ series power supplies are designed for use in medical, information technology or laboratory  environments. They provide a "universal"  AC input operation, and can supply 0 to approximately 50W of stable DC output power.

1.1 The model number suffix "M", indicates that a specific model is for use in a medical environment.

2. Electrical Specifications:

2.1 Input: 100-250Vac     50/60Hz    120VA
(90-264 Vac, 47-63 Hz is the tested operational range, 100-250 Vac range is marked on product to comply with safety standards.)

2.1.1 Input Fuses:
A current fuse is incorporated in the AC input line, rated at 3A, 250V, slo blo type fuse element, 2AG size, configured with integral welded on leads. Available from Littlefuse, part number 230-3. Medical grade models have a second fuse in the AC input neutral line as well.

2.1.2 Input Switch:
An optional input switch may be provided if required. A "I" and a "O" will be marked on it, or next to it.

2.1.3 Insulation Class: Class I equipment.

2.2 Output:

  Optional suffix "M", indicates medical usage. Optional suffix "Q" indicates quasi regulated.

 Single Output Models
 

WSZ150 
WSZ155 
WSZ160 
WSZ165 
WSZ170 
WSZ175 
WSZ180 
WSZ190 
WSZ110 
WSZ111 
WSZ112 
WSZ113 
WSZ114 
WSZ115 

5.0Vdc 8A
5.5Vdc 7.3A
6.0Vdc 6.7A
6.5Vdc 6.2A
7.0Vdc 5.7A
7.5Vdc 5.3A
8.0Vdc 5A
9.0Vdc 4.8A
10Vdc 4.3A
11Vdc 4A
12Vdc 4.2A
13Vdc 3.8A
14Vdc 3.6A
15Vdc 3.3A

WSZ116 
WSZ118 
WSZ120 
WSZ122 
WSZ124 
WSZ126 
WSZ128 
WSZ130 
WSZ132 
WSZ136 
WSZ140 
WSZ144 
WSZ148
WSZ191  

16Vdc 3.1A
18Vdc 2.8A
20Vdc 2.5A
22Vdc 2.3A
24Vdc 2.1A
26Vdc 1.9A
28Vdc 1.8A
30Vdc 1.7A
32Vdc 1.6A
36Vdc 1.4A
40Vdc 1.3A
44Vdc 1.2A
48Vdc 1.1A
12Vdc 4.2A

 Dual Output Models
 

WSZ201 
alternate rating

WSZ202 
alternate rating

WSZ203 
alternate rating

WSZ204 
alternate rating

WSZ205 
WSZ206 
WSZ207 

WSZ208 
alternate rating

WSZ209 
alternate rating

WSZ210 
alternate rating

+12Vdc  2A
+12Vdc  3A

+15Vdc  1.6A
+15Vdc  2.2A

+20Vdc  1.2A
+20Vdc  1.4A

+24Vdc  1A
+24Vdc  1.5A

+5Vdc  7A
+5Vdc  4A
+5Vdc  4A

+5Vdc  6.5A
+5Vdc  5A

+5Vdc  6.5A
+5Vdc  4A

+12Vdc  2.4A
+12Vdc  2A

-12Vdc 2A
-12Vdc 1A

-15Vdc 1.6A
-15Vdc 1A

-20Vdc 1.2A
-20Vdc 1A

-24Vdc 1A
-24Vdc .5A

-5Vdc  2A
+12Vdc  2A
+15Vdc  1.6A

+20Vdc  .5A
+20Vdc  1A

+24Vdc  .5A
+24Vdc  1A

+18Vdc  1A
+18Vdc  1.25A

  Triple Output Models
 

WSZ301 
alternate rating
alternate rating
alternate rating

WSZ302 
alternate rating
alternate rating
alternate rating

+5Vdc  4A
+5Vdc  5A
+5Vdc  6A
+5Vdc  6A

+5Vdc  3A
+5Vdc  5A
+5Vdc  6A
+5Vdc  6A

+12Vdc  1A
+12Vdc  1A
+12Vdc  .6A
+12Vdc  1A

+15Vdc  1A
+15Vdc  1A
+15Vdc  .5A
+15Vdc  .75A

-12Vdc 1A
-12Vdc .6A
-12Vdc .6A
-12Vdc .25A

-15Vdc 1A
-15Vdc .3A
-15Vdc .5A
-15Vdc .25A

 2.3 Output Variance, Voltage:
+/- 5% at Full Load, any input voltage or environmental condition, for single output models, multi output models may be specified differently, contact factory.

2.4 Load Regulation, Voltage: (either option A or B may be provided)
A)  Fully Regulated Models: Not to exceed 600 millivolts, for 100% to 0% load change, varies due to output cord selected, regulation at PCB better than 1%. Optocoupler used to feed back control signal to primary controller.

B)  Quasi Regulated Models: 10% maximum increase for a 100% to 20% load change. 30% maximum increase for a 100% to 0% load change. Primary side voltage regulation technique used.

2.5 Protection Circuitry: The output is protected from short circuit and prolonged overload.

2.6 Indicators: A optional green LED may be provided in the output circuit if required.

3. Mechanical Specifications:

3.1 Enclosure Dimensions: 6.50"L X 3.75"W X 2.17"H inches, enclosure outline drawings available upon request.

3.2 Input Configuration: Male IEC320, 3 conductor, power inlet.

3.3 Output Configuration:   6 foot length standard, various conductor arrangements and jacket material selections are available. Cordsets may be terminated with various dc connectors, custom output connector assemblies are available, contact customer service for details. Standard in-stock cordsets can be seen in the Jerome Industries engineering manual / catalog on pages 10 and 11.

4. Environmental:
 

4.1 Operating :

Temperature: 0 to 40C
Humidity: 10% to 95% non-condensing
Atmospheric pressure: 70 to 106 kPa

Please contact customer service if a
wider operating range is required.

 4.2 Storage :

Temperature: -40 to 75C
Humidity: 10% to 95% non-condensing
Atmospheric pressure: 50 to 106 kPa

4.3 Ingress Protection:  Standard IPX0,
 IPX1 optionally available upon request.

5. Regulatory Compliance

5.1 Safety Compliance:
These power supplies have been certified by the following safety agencies to the standards shown:
 

Regulatory Agency

UL
 
 

UL testing and 
evaluation of 
Canadian 
requirements.

TUV
 

Standard

UL60601-1
UL544 
UL60950-1 

C22.2#601-1
C22.2#125 
C22.2#60950-1

EN60601-1
EN60950-1 

Classification

Isolation, BF
Patient Care, Grounded
SELV

Isolation, BF
Risk Class 2G
SELV

Isolation, BF
SELV

File#

E160908
E160908
E148586

E160908
E160908
E148586

B030924072012
B030924072012

5.3 Electro Magnetic Compatibility:
Meets requirements of European EMC directive and FCC requirements.

5.3.1 Medical:
Meets EMC requirements from EN60601-1-2 (EMC requirements for ElectroMedical Equipment).

5.3.2 Emissions:
Meets level B requirements of FCC 47 CFR part 15 and CISPR11 (EN55011) and CISPR22 (EN55022), EN61000-6-3 (light industrial) and line harmonic emissions EN61000-3-2.

5.3.3 Immunity:
Meets CISPR24 (EN55024), EN61326 and EN61000-6-1.

5.3.4 EMC Responsibility:
It is the responsibility of the OEM customer to verify EMC compliance with his system before marketing the end-use system.
 

 Notes:

1) Emissions have been verified by an independent lab on 4 models of the WSZ family. The models tested were the 5V, 12V and 48V single output and the triple output WSZ301 model. All models were found to comply with radiated and conducted emissions requirements.
Emissions are monitored on all prototypes using a spectrum analyzer with peak, quasi peak and average mode power measurement capabilities. Conducted emissions are recorded for both 120 Vac and 230 Vac input conditions. A correlated radiated emissions program is used in order to economically verify the radiated emissions from a specific model.
The radiated emissions program works from the premise that the power supply acts as a dipole antenna radiator, one side of the dipole antenna being the input cord, the other side being the output cord. Levels are monitored on the cords by a common mode current sensing device operating into a spectrum analyzer. Correlation factors have been established which are based upon the difference between radiated field measurements made at an independent lab and Jerome Industries lab common mode current measurements on the cords of reference power supplies.
Using this technique, Jerome Industries has experienced a 98% passing rate for prototypes submitted for radiated field emissions testing at independent labs.

OEM System Engineer Instructions

*  The EMC characteristics of the final system comprised of both the power supply and the main system must be evaluated to the applicable EMC environment regulations.

*  Should a power supply need repair, please return it along with a note describing the particular problem to:

  Jerome Industries Inc.
  730 Division St.
  Elizabeth, NJ 07201
  U.S.A.

*  Circuit operation block diagrams, isolation diagrams, and a list of safety critical components will be available on request to appropriately qualified personnel.

* All products which contain double input fuses shall contain the following text in the system installation manual:
"CAUTION: DOUBLE POLE/NEUTRAL FUSING"

Medical Products

*  When used with medical products the Instruction Manual prepared for the end use product must incorporate at a minimum, the following power supply warning notices:
 

 a) Warning: When cleaning the power unit, use a cloth dampened with cleaning alcohol on the outside   of the enclosure only. Do not immerse product in water or a safety hazard could arise during use.
 b) Warning: Do not use this power unit in the presence of flammable anesthetics.

* The Instruction manual must also contain a notice regarding methods of disposal of the equipment at the end it's useful life.

*  Periodic leakage current testing should be done on the combined power supply and end-use system on a yearly basis when used in a hospital environment where such test equipment is commonly available.

* Per IEC601-1 clause 56.8, an indication must be provided to the operator that the end-use equipment is functioning as intended. A green LED and/or a power switch can be provided on the power supply if required for your specific system.


Appendix A
WSZ100 Family Leakage Current
Measurements per IEC60601-1 Clause 19

  NC SFC
Earth Leakage Current 180.93 344.7
Enclosure Leakage Current 1.09 1.62
Patient Leakage Current
(to output negative cord connection)
7.87 158.83
Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) N/A 2363

Notes:
1) Measurement data in microamps
2) Measurements made with 275 Vac, 60 Hz input.
3) Worst case measurement condition, including humidity preconditioning.
4) Reduced leakage designs available (including CF category compliant designs), contact factory for availability.

 