WSA100 FAMILY SWITCHMODE POWER SUPPLY SPECIFICATIONS

Specifications:

1. General Character:
The model WSA series power supplies are designed for use in medical, information technology or laboratory  environments. They provide a "universal"  AC input operation, and can supply 0 to approximately 15W of stable DC output power.

1.1 These units are available in various international wall plug-in configurations, or with an IEC320 power inlet for easy universal input configurability.

1.2 The model number suffix "M", indicates that a specific model is for use in a medical environment.

1.3 The model number suffix "Q", indicates that a specific model has a quasi-regulated output characteristic.

1.4 Optional suffix indicates mechanical input configuration:

  N, North American (or NEMA) plug-in
  U, United Kingdom plug-in BS1363
  E, European mainland, CEE7/16
  A, Australian, AS3112
  If blank, then the standard IEC320 inlet is specified.

2. Electrical Specifications:

2.1 Input: 100-240Vac     50/60Hz     40VA
(90-264 Vac, 47-63 Hz is the tested operational range, 100-240 Vac range is marked on product to comply with safety standards.)

2.1.1 Input Fuses:
A current fuse is incorporated in the AC input line, rated at 1A, 250V, slo blo type fuse element, 2AG size, configured with integral welded on leads. Available from Littlefuse, part number 230-1. Medical grade models supplied with a 3 conductor input configuration have a second fuse in the AC input neutral line as well.

2.1.3 Protection Class: Class II equipment (with optional functional earth ground)

2.2 Output:
 

WSA150 
WSA155 
WSA160 
WSA165 
WSA170 
WSA175 
WSA180 
WSA190 
WSA110 
WSA111 
WSA112 
WSA113 
WSA114 
WSA115 

5.0Vdc 2.5A
5.5Vdc 2.25A
6.0Vdc 2.07A
6.5Vdc 1.9A
7.0Vdc 1.75A
7.5Vdc 1.65A
8.0Vdc 1.55A
9.0Vdc 1.5A
10Vdc 1.35A
11Vdc 1.25A
12Vdc 1.25A
13Vdc 1.15A
14Vdc 1.07A
15Vdc 1.0A

WSA116
WSA118 
WSA120 
WSA122 
WSA124 
WSA126 
WSA128 
WSA130 
WSA132 
WSA136 
WSA140 
WSA144 
WSA148

16Vdc .925A
18Vdc .825A
20Vdc .75A
22Vdc .675A
24Vdc .625A
26Vdc .575A
28Vdc .535A
30Vdc .50A
32Vdc .50A
36Vdc .45A
40Vdc .40A
44Vdc .375A
48Vdc .35A

2.3 Output Variance, Voltage:
+/- 5% at Full Load, any input voltage or environmental condition, for standard output cordsets.

2.4 Load Regulation, Voltage: (either option A or B may be provided)
A)  Fully Regulated Models: Output will stay in +/- 5% window from 100% to no load. Optocoupler used to feed back control signal to primary controller.

B)  Quasi Regulated Models: 20% maximum increase for a 100% to 20% load change. 30% maximum increase for a 100% to 0% load change. Primary side voltage regulation technique used.

2.5 Protection Circuitry: The output is protected from short circuit and prolonged overload.

3. Mechanical Specifications:

3.1 Enclosure Dimensions: 3.25"L X 2.72"W X 1.89"H inches, enclosure outline drawings available upon request.
Standard color black, custom colors available, consult factory.

3.2 Input Configuration:

Option 1) Male IEC320, 3 conductor, power inlet. Additional detachable input cordset must be supplied in   addition to power adapter. Standard international configuration, 100-240V rated.
Option 2) Wall Plug-in NEMA 5-15P or 1-15P, Standard North American style for 120V, also used in    Japan and some other area's in the far east. 100-120V rated.
Option 3) Wall Plug-in BS1363 U.K. prong assembly, plastic ground pin. 220-240V rated.
Option 4) Wall Plug-in CEE7/16 standard European mainland 2 conductor. 220-240V rated.
Option 5) Wall Plug-in AS3112 standard Australian 2 or 3 conductor. 220-240V rated.

3.3 Output Configuration:   6 foot length standard, various conductor arrangements and jacket material selections are available. Cordsets may be terminated with various dc connectors, custom output connector assemblies are available, contact customer service for details. Standard in-stock cordsets can be seen in the Jerome Industries engineering manual / catalog on pages 10 and 11.

4. Environmental:
 

4.1 Operating : 

Temperature: 0 to 40C
Humidity: 10% to 95% non-condensing
Atmospheric pressure: 70 to 106 kPa

Please contact customer service if a
wider operating range is required.

4.2 Storage :

 Temperature: -40 to 75C
 Humidity: 10% to 95% non-condensing
 Atmosperic pressure: 50 to 106 kPa

5. Regulatory Compliance

5.1 Safety Compliance:
These power supplies have been certified by the following safety agencies to the standards shown:
 

Regulatory Agency

UL
 
 

UL testing and 
evaluation of 
Canadian 
requirements.

TUV
 

Standard

UL60601-1
UL544 
UL60950-1 

C22.2#601-1
C22.2#125 
C22.2#60950-1
 

EN60601-1
EN60950-1

Classification

Isolation, BF
Patient Care, Grounded
SELV

Isolation, BF
Risk Class 2G
SELV
 

Isolation, BF
SELV

File#

E160908
E160908
E148586

E160908
E160908
E148586
 

B971024072007
B971024072007

5.3 Electro Magnetic Compatibility:
Meets requirements of European EMC directive and FCC requirements.

5.3.1 Medical:
Meets EMC requirements from EN60601-1-2 (EMC requirements for ElectroMedical Equipment).

5.3.2 Emissions:
Meets level B requirements of FCC 47 CFR part 15 and CISPR11 (EN55011) and CISPR22 (EN55022), EN61000-6-3 (light industrial) and line harmonic emissions EN61000-3-2.

5.3.3 Immunity:
Meets CISPR24 (EN55024), EN61326 and EN61000-6-1.

5.3.4 EMC Responsibility:
It is the responsibility of the OEM customer to verify EMC compliance with his system before marketing the end-use system.
 

 Notes:

1) Emissions are monitored on all prototypes using a spectrum analyzer with peak, quasi peak and average mode power measurement capabilities. Conducted emissions are recorded for both 120 Vac and 230 Vac input conditions. A correlated radiated emissions program is used in order to economically verify the radiated emissions from a specific model.
The radiated emissions program works from the premise that the power supply acts as a dipole antenna radiator, one side of the dipole antenna being the input cord, the other side being the output cord. Levels are monitored on the cords by a common mode current sensing device operating into a spectrum analyzer. Correlation factors have been established which are based upon the difference between radiated field measurements made at an independent lab and Jerome Industries lab common mode current measurements on the cords of reference power supplies.
Using this technique, Jerome Industries has experienced a 98% passing rate for prototypes submitted for radiated field emissions testing at independent labs.

OEM System Engineer Instructions

*  The EMC characteristics of the final system comprised of both the power supply and the main system must be evaluated to the applicable EMC environment regulations.

*  Should a power supply need repair, please return it along with a note describing the particular problem to:

Jerome Industries Inc.
730 Division St.
Elizabeth, NJ 07201
U.S.A.

*  Circuit operation block diagrams, isolation diagrams, and a list of safety critical components will be available on request to appropriately qualified personnel.

* All products which contain double input fuses shall contain the following text in the system installation manual:
"CAUTION: DOUBLE POLE/NEUTRAL FUSING"

Medical Products

*  When used with medical products the Instruction Manual prepared for the end use product must incorporate at a minimum, the following power supply warning notices:
 

 a) Warning: When cleaning the power unit, use a cloth dampened with cleaning alcohol on the outside of the enclosure only. Do not immerse product in water or a safety hazard could arise during use.
 b) Warning: Do not use this power unit in the presence of flammable anesthetics.

* The Instruction manual must also contain a notice regarding methods of disposal of the equipment at the end it's useful life.

*  Periodic leakage current testing should be done on the combined power supply and end-use system on a yearly basis when used in a hospital environment where such test equipment is commonly available.

*  This product is classified as Class 2 (double insulated), the optional input functional ground is used only for reduction of leakage current available at the output connections.


Appendix A
WSA100 Family Leakage Current
Measurements per IEC60601-1 Clause 19

CLASS 1 CONFIGURATION
  NC SFC
Earth Leakage Current 187.1 283.5
Enclosure Leakage Current 1.34 2.18
Patient Leakage Current
(to output negative cord connection)
21.6 94.44
Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) N/A 325.2

 

CLASS 2 CONFIGURATION
  NC SFC
Enclosure Leakage Current 1.58 2.66
Patient Leakage Current
(to output negative cord connection)
50.33 74.71
Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) N/A 131.19

Notes:
1) Measurement data in microamps
2) Measurements made with 275 Vac, 60 Hz input.
3) Worst case measurement condition, including humidity preconditioning.
4) Reduced leakage designs available (including CF category compliant designs), contact factory for availability.

 

Back Home