WSZ700/800/900 FAMILY SWITCHMODE POWER SUPPLY SPECIFICATIONS

Specifications:

1. General Character:
The model WSZ series power supplies are designed for use in medical, information technology or laboratory  environments. They provide a "universal"  AC input operation, and can supply 0 to approximately 80W of stable DC output power during cycled loading. The continuous loading limit is approximately 60W.

1.1 The model number suffix "M", indicates that a specific model is for use in a medical environment.

2. Electrical Specifications:

2.1 Input: 100-250Vac     50/60Hz    
(90-264 Vac, 47-63 Hz is the tested operational range, 100-250 Vac range is marked on product to comply with safety standards.)

2.1.1 Input Fuses:
A current fuse is incorporated in the AC input line, rated at 3A, 250V, slo blo type fuse element, 2AG size, configured with integral welded on leads. Available from Littlefuse, part number 230-3. Medical grade models have a second fuse in the AC input neutral line as well.

2.1.2 Input Switch:
An optional input switch may be provided if required. A "I" and a "O" will be marked on it, or next to it.

2.2 Output:

   Single Output Models
 

 
Input (ac)
Output (dc)
Model Continuous
VA
Surge
VA
Volts Continuous
Amps
Continuous
Watts
Surge
Amps
Surge
Watts

WSZ750
WSZ755
WSZ760
WSZ765
WSZ770
WSZ775
WSZ780
WSZ790
WSZ710
WSZ711
WSZ712
WSZ713
WSZ714
WSZ715
WSZ716
WSZ718
WSZ720
WSZ722
WSZ724
WSZ726
WSZ728
WSZ730
WSZ732
WSZ736
WSZ740
WSZ744
WSZ748

130
130
130
130
130
130
130
130
150
150
150
150
150
150
150
150
150
150
150
150
150
150
150
150
150
150
150

170
170
170
170
170
170
170
170
200
200
200
200
200
200
200
200
200
200
200
200
200
200
200
200
200
200
200

5.0
5.5
6.0
6.5
7.0
7.5
8.0
9.0
10
11
12
13
14
15
16
18
20
22
24
26
28
30
32
36
40
44
48

10.00
9.10
8.30
7.70
7.10
6.7
0
6.30
5.50
5.00
5.00
5.00
4.60
4.30
4.00
3.75
3.30
3.00
2.70
2.50
2.30
2.10
2.00
1.90
1.70
1.50
1.40
1.25

50
50
50
50
50
50
50
50
50
55
60
60
60
60
60
60
60
60
60
60
60
60
60
60
60
60
60

12.00
10.90
10.80
10.00
9.30
8.70
8.10
7.20
6.70
6.70
6.70
6.20
5.70
5.30
5.00
4.40
4.00
3.60
3.30
3.10
2.90
2.70
2.50
2.20
2.00
1.85
1.70

60
60
65
65
65
65
65
65
67
74
80
80
80
80
80
80
80
80
80
80
80
80
80
80
80
80
80

 Multi Output Models

 
Input (ac)
Output (dc)
Model Continuous
VA
Surge
VA
Volts Continuous
Amps
Continuous
Watts
Surge
Amps
Surge
Watts

WSZ801

WSZ802

WSZ803

WSZ804


WSZ805

WSZ806

WSZ807

WSZ808


WSZ809

WSZ810

WSZ901


WSZ902

150

150

150

150


130

130

130

130


150

150

130


130

200

200

200

200


170

170

170

170


200

200

170


170

+12
   -12
+15
   -15
+20
   -20
+24
   -24

+5
   -5
+5
   +12
+5
   +15
+5
   +24

+12
   +18
+12
   +24
+5
   +12
   -12
+5
   +15
   -15

5.0
   2.5
4.0
   2.0
3.0
   2.0
2.5
   1.5

8.0
   4.0
8.0
   3.0
8.0
   3.0
8.0
   2.0

5.0
   3.0
5.0
   2.0
8.0
   2.0
   2.0
8.0
   1.6
   1.6

60

60

60

60


50

50

50

50


60

60

50


50

6.7
   3.5
5.3
   3.0
4.0
   3.0
3.3
   2.5

12.0
   6.0
12.0
   4.0
12.0
   4.0
12.0
   2.5

6.7
   4.0
6.7
   2.5
12.0
   3.0
   3.0
12.0
   2.4
   2.4

80

80

80

80


60

65

65

65


80

80

65


65

      Programmable Output Models
 
 
Input (ac)
Output (dc)
Model Continuous
VA
Surge
VA
Volts Continuous
Amps
Continuous
Watts
Surge
Amps
Surge
Watts

WSZ791

WSZ792

WSZ793

WSZ794

WSZ795

WSZ796

WSZ895

130

130

150

150

150

150

150

170

170

200

200

200

200

200

5.0 to 6.5
   
7.0 to 9.0
   
10 to 16
   
18 to 30
   
32 to 48
 
5 to 30
   
5 to 30
+12

10 to 7.7
   
7.1 to 5.5
   
5.0 to 3.75
   
3.3 to 2.0
   
1.9 to 1.25
 
3.3 to 1.7
   
3.3 to 1.7
0.80

50

50

60

60

60

17-50

17-50

12 to 9.2
   
8.6 to 6.7
   
6.7 to 5.0
   
4.4 to 2.7
   
2.5 to 1.7
   
3.0 to 1.5
   
5.0 to 2.5
0.80

60

65

80

80

80

25-75

25-75

2.3 Output Variance, Voltage:
+/- 5% at Full Load, any input voltage or environmental condition, for single output models. Multi output models may be specified differently, contact factory.

2.4 Load Regulation, Voltage:
Not to exceed 600 millivolts, for 100% to 0% load change, varies due to output cord selected, regulation at PCB better than 1%.

2.5 Protection Circuitry: The output is protected from short circuit and prolonged overload.

3. Mechanical Specifications:

3.1 Enclosure Dimensions: 6.50"L X 3.75"W X 2.17"H inches, enclosure outline drawings available upon request.

3.2 Input Configuration:

Option 1) Male IEC320, 3 conductor, power inlet. Additional detachable input cordset must be supplied in   addition to power adapter. Standard international configuration, 100-240V rated.
Option 2) Fixed-in-place Input Cord, contact factory for availability.

3.3 Output Configuration:   6 foot length standard, various conductor arrangements and jacket material selections are available. Cordsets may be terminated with various dc connectors, custom output connector assemblies are available, contact customer service for details. Standard in-stock cordsets can be seen in the Jerome Industries engineering manual / catalog on pages 10 and 11.

4. Environmental:

4.1 Operating :
Temperature: 0 to 40C
Humidity: 10% to 95% non-condensing
Atmospheric pressure: 70 to 106 kPa
Please contact customer service if a
wider operating range is required.

4.2 Storage :

Temperature: -40 to 75C
Humidity: 10% to 95% non-condensing
Atmosperic pressure: 50 to 106 kPa

5. Regulatory Compliance

5.1 Safety Compliance:  These Power Supplies have been certified by the following safety agencies to the standards shown:
 

Regulatory Agency

UL

 
 UL testing and 
evaluation of 
Canadian 
requirements.

TUV
 

Standard

UL60601-1
UL60950-1 

C22.2#601-1
C22.2#60950-1

 

EN60601-1
EN60950-1 

Classification

Isolation, BF
SELV

Isolation, BF
SELV
 
 

Isolation, BF
SELV

File#

E160908
E148586

E160908
E148586

 

B030924072012
B030924072012

5.3 Electro Magnetic Compatibility:
Meets requirements of European EMC directive and FCC requirements.

5.3.1 Medical:
Meets EMC requirements from EN60601-1-2 (EMC requirements for ElectroMedical Equipment).

5.3.2 Emissions:
Meets level B requirements of FCC 47 CFR part 15 and CISPR11 (EN55011) and CISPR22 (EN55022), EN61000-6-3 (light industrial) and line harmonic emissions EN61000-3-2.

5.3.3 Immunity:
Meets CISPR24 (EN55024), EN61326 and EN61000-6-1.

5.3.4 EMC Responsibility:
It is the responsibility of the OEM customer to verify EMC compliance with his system before marketing the end-use system.
 

Notes:

1) Emissions are monitored on all prototypes using a spectrum analyzer with peak, quasi peak and average mode power measurement capabilities. Conducted emissions are recorded for both 120 Vac and 230 Vac input conditions. A correlated radiated emissions program is used in order to economically verify the radiated emissions from a specific model.
The radiated emissions program works from the premise that the power supply acts as a dipole antenna radiator, one side of the dipole antenna being the input cord, the other side being the output cord. Levels are monitored on the cords by a common mode current sensing device operating into a spectrum analyzer. Correlation factors have been established which are based upon the difference between radiated field measurements made at an independent lab and Jerome Industries lab common mode current measurements on the cords of reference power supplies.
Using this technique, Jerome Industries has experienced a 98% passing rate for prototypes submitted for radiated field emissions testing at independent labs.

OEM System Engineer Instructions

*  The EMC characteristics of the final system comprised of both the power supply and the main system must be evaluated to the applicable EMC environment regulations.

*  Should a power supply need repair, please return it along with a note describing the particular problem to:

  Jerome Industries Inc.
  730 Division St.
  Elizabeth, NJ 07201
  U.S.A.

*  Circuit operation block diagrams, isolation diagrams, and a list of safety critical components will be available on request to appropriately qualified personnel.

* All products which contain double input fuses shall contain the following text in the system installation manual:
"CAUTION: DOUBLE POLE/NEUTRAL FUSING"

Medical Products

*  When used with medical products the Instruction Manual prepared for the end use product must incorporate at a minimum, the following power supply warning notices:
 

 a) Warning: When cleaning the power unit, use a cloth dampened with cleaning alcohol on the outside   of the enclosure only. Do not immerse product in water or a safety hazard could arise during use.
 b) Warning: Do not use this power unit in the presence of flammable anesthetics.
 

* The Instruction manual must also contain a notice regarding methods of disposal of the equipment at the end it's useful life.

*  Periodic leakage current testing should be done on the combined power supply and end-use system on a yearly basis when used in a hospital environment where such test equipment is commonly available.


Appendix A
WSZ700 Family Leakage Current
Measurements per IEC60601-1 Clause 19

  NC SFC
Earth Leakage Current 223.4 377.7
Enclosure Leakage Current 2.1 3.85
Patient Leakage Current
(to output negative cord connection)
86.3 140.5
Patient Leakage Current, Mains Voltage Applied (to output negative cord connection) N/A 407.9

Notes:
1) Measurement data in microamps
2) Measurements made with 275 Vac, 60 Hz input.
3) Worst case measurement condition, including humidity preconditioning.
4) Reduced leakage designs available (including CF category compliant designs), contact factory for availability.