EXCEL FAMILY SWITCHMODE POWER SUPPLY SPECIFICATIONS

Specifications:

1. General Character:
The model WSL400, WSL500, WSX400, WSZ400 and WSZ500 series power supplies are designed for use in medical, information technology or laboratory  environments. They provide a "universal"  AC input operation.

1.1 The model number suffix "M", indicates that a specific model is for use in a medical environment.

1.2 Optional suffix indicates mechanical input configuration:

  N, North American (or NEMA) plug-in
  U, United Kingdom plug-in BS1363
  E, European mainland, CEE7/16
  A, Australian, AS3112
  H, Non-detachable Hospital Grade input cord
  If blank, then the standard IEC320 inlet is specified.

2. Electrical Specifications:

2.1 Input: 100-250Vac     50/60Hz
(90-264 Vac, 47-63 Hz is the tested operational range, 100-250 Vac range is marked on product to comply with safety standards.)

2.1.1 Input Fuses:
A current fuse is incorporated in each of the 2 AC lines, slo blo type fuse element, 2AG size, configured with integral welded on leads. Fuse Rating: WSL400 family = 1.25A, 250V / WSL500 family = 2A, 250V / WSX400, WSZ400 and WSZ500 families = 3.5A, 250V.

2.1.2 Input Switch:
An optional input switch may be provided if required. A "I" and a "O" will be marked on it, or next to it.

2.2 Output:  
( Optional model number suffix letter "M", indicates medical usage. )

WSL433
WSL439
WSL445
WSL450
WSL455
WSL460
WSL465
WSL470
WSL475
WSL480
WSL490
WSL410
WSL411
WSL412 

WSL413 
WSL414 
WSL415
WSL416 

WSL418 
WSL420 
WSL422
WSL424 

WSL426 
WSL428 
WSL430 
WSL432 
WSL436 
WSL440 
WSL444 
WSL448 
3.3Vdc 8.2A
3.9Vdc 8.2A
4.5Vdc 8.2A
5.0Vdc 8.2A
5.5Vdc 7.6A
6.0Vdc 7.1A
6.5Vdc 6.7A
7.0Vdc 6.2A
7.5Vdc 5.83A
8.0Vdc 5.5A
9.0Vdc 5.0A
10Vdc 4.7A
11Vdc 4.4A
12Vdc 4.2A

13Vdc 3.85A 
14Vdc 3.55A
15Vdc 3.3A
16Vdc 3.1A
18Vdc 2.8A
20Vdc 2.5A
22Vdc 2.3A
24Vdc 2.1A
26Vdc 1.9A
28Vdc 1.8A
30Vdc 1.7A
32Vdc 1.6A
36Vdc 1.4A
40Vdc 1.25A
44Vdc 1.14A
48Vdc 1.04A
  WSL533
WSL539
WSL545
WSL550
WSL555
WSL560
WSL565
WSL570
WSL575
WSL580
WSL590
WSL510
WSL511
WSL512 

WSL513 
WSL514 
WSL515
WSL516 

WSL518 
WSL520 
WSL522   
WSL524 
WSL526 
WSL528 
WSL530 
WSL532 
WSL536 
WSL540 
WSL544 
WSL548 

3.3Vdc 10A
3.9Vdc 10A
4.5Vdc 10A
5.0Vdc 10A
5.5Vdc 9.4A
6.0Vdc 8.6A
6.5Vdc 8.0A
7.0Vdc 7.4A
7.5Vdc 6.9A
8.0Vdc 6.5A
9.0Vdc 6.0A
10Vdc 5.5A
11Vdc 5.0A

12Vdc 5.0A
13Vdc 4.6A
14Vdc 4.3A
15Vdc 4.0A
16Vdc 3.75A
18Vdc 3.3A
20Vdc 3.0A
22Vdc 2.7A
24Vdc 2.5A
26Vdc 2.3A
28Vdc 2.15A
30Vdc 2.0A
32Vdc 1.85A
36Vdc 1.65A
40Vdc 1.50A
44Vdc 1.35A
48Vdc 1.25A

  WSZ433
WSZ439
WSZ445
WSZ450
WSZ455
WSZ460
WSZ465
WSZ470
WSZ475
WSZ480
WSZ490
WSZ410
WSZ411
WSZ412 

WSZ413 
WSZ414 
WSZ415
WSZ416 

WSZ418 
WSZ420 
WSZ422 
WSZ424 

WSZ426 
WSZ428 
WSZ430 
WSZ432 
WSZ436 
WSZ440 
WSZ444 
WSZ448     
3.3Vdc 12A
3.9Vdc 12A
4.5Vdc 12A
5.0Vdc 12A
5.5Vdc 10.9A
6.0Vdc 10.8A
6.5Vdc 10A
7.0Vdc 9.3A
7.5Vdc 8.7A
8.0Vdc 8.1A
9.0Vdc 7.2A
10Vdc 6.7A
11Vdc 6.7A
12Vdc 6.7A

13Vdc 6.2A 
14Vdc 5.7A
15Vdc 5.3A
16Vdc 5.0A
18Vdc 4.5A
20Vdc 4.0A
22Vdc 3.6A
24Vdc 3.3A

26Vdc 3.1A
28Vdc 2.9A
30Vdc 2.7A
32Vdc 2.5A
36Vdc 2.2A
40Vdc 2.0A
44Vdc 1.8A
48Vdc 1.7A  

 

WSZ512 
WSZ513 
WSZ514 
WSZ515
WSZ516 

WSZ518 
WSZ520 
WSZ522
WSZ524 

WSZ526 
WSZ528 
WSZ530 
WSZ532 
WSZ536 
WSZ540 
WSZ544 
WSZ548 
12Vdc 7.5A
13Vdc 6.9A 
14Vdc 6.4A
15Vdc 6.0A
16Vdc 5.6A
18Vdc 5.0A
20Vdc 4.5A
22Vdc 4.1A
24Vdc 3.8A
26Vdc 3.5A
28Vdc 3.2A
30Vdc 3.0A
32Vdc 2.8A
36Vdc 2.5A
40Vdc 2.3A
44Vdc 2.0A
48Vdc 1.9A
  WSX412 
WSX413 
WSX414 
WSX415
WSX416 

WSX418 
WSX420 
WSX422   
WSX424 
WSX426 
WSX428 
WSX430 
WSX432 
WSX436 
WSX440 
WSX444 
WSX448 
12Vdc 8.3A
13Vdc 7.7A 
14Vdc 7.15A
15Vdc 6.7A
16Vdc 6.25A
18Vdc 5.55A
20Vdc 5.0A
22Vdc 4.5A
24Vdc 4.2A
26Vdc 3.8A
28Vdc 3.6A
30Vdc 3.3A
32Vdc 3.1A
36Vdc 2.8A
40Vdc 2.5A
44Vdc 2.3A
48Vdc 2.1A
     

2.3 Output Variance, Voltage:
+/- 5% at Full Load, any input voltage or environmental condition

2.4 Load Regulation, Voltage:
Not to exceed 800 millivolts for 100% to 0% load change, varies due to output cord selected, remote sense option available, contact factory.

2.5 Protection Circuitry: The output is protected from short circuit and prolonged overload.

3. Mechanical Specifications:

3.1 Enclosure Dimensions:
WSL400 and WSL500 families:4.75"L X 3.00"W X 1.90 "H inches. WSZ400 and WSZ500 families: 6.5 "L X 3.75"W X 2.17"H inches. WSX400 family: 6.5 "L X 3.75"W X 2.56"H inches. Enclosure outline drawings available upon request.

3.2 Input Configuration: Male IEC320 power inlet, 2 or 3 conductor option. Optional input connector lock available. Optional non-detachable input cords available for various international requirements.

3.3 Output Configuration:   6 foot length standard, various conductor arrangements and jacket material selections are available. Cordsets may be terminated with various dc connectors, custom output connector assemblies are available, contact customer service for details. Standard in-stock cordsets can be seen in the Jerome Industries engineering manual / catalog on pages 10 and 11.

4. Environmental:
 

4.1 Operating :

Temperature: 0 to 40C 
Humidity: 10% to 95% non-condensing
Atmospheric pressure: 70 to 106 kPa

Please contact customer service if a
wider operating range is required.

4.2 Storage :

Temperature: -40 to 75C
Humidity: 10% to 95% non-condensing
Atmosperic pressure: 50 to 106 kPa

5. Regulatory Compliance

5.1 Safety Compliance:
These power supplies have been certified by the following safety agencies to the standards shown:
 
5.1.1 Single Output Models
These power supplies have been certified by the following safety agencies to the standards shown:

Regulatory Agency

UL
 
 UL testing and 
evaluation of 
Canadian 
requirements.  

 

TUV
 

Standard

UL60601-1
UL60950-1 


C22.2#601-1

C22.2#60950-1



EN60601-1

EN60950-1 

Classification

Isolation, BF
SELV


Isolation, BF

SELV



Isolation, BF

SELV

File#

E160908
E148586


E160908

E148586



B040924072017

B040924072017

5.2 Power supplies shipped to European market need to bear the CE mark. Jerome Industries can supply a CE mark under the following standards:
 
Standard

EN60601-1
EN60950-1
Classification

Isolation, BF
SELV

5.3 Electro Magnetic Compatibility:
Meets requirements of European EMC directive and FCC requirements.

5.3.1 Medical:
Meets EMC requirements from EN60601-1-2 (EMC requirements for ElectroMedical Equipment).

5.3.2 Emissions:
Meets level B requirements of FCC 47 CFR part 15 and CISPR11 (EN55011) and CISPR22 (EN55022), EN61000-6-3 (light industrial) and line harmonic emissions EN61000-3-1.

5.3.3 Immunity:
Meets CISPR24 (EN55024), EN61326 and EN61000-6-3.

5.3.4 EMC Responsibility:
It is the responsibility of the OEM customer to verify EMC compliance with his system before marketing the end-use system.
 

 Notes:

1) Emissions are monitored on all prototypes using a spectrum analyzer with peak, quasi peak and average mode power measurement capabilities. Conducted emissions are recorded for both 120 Vac and 230 Vac input conditions. A correlated radiated emissions program is used in order to economically verify the radiated emissions from a specific model.
The emissions program works from the premises of the power supply acting as a dipole antenna radiator, one side of the dipole antenna being the input cord, the other side being the output cord. Levels are monitored on the cords by a common mode current sensing device operating into a spectrum analyzer. Correlation factors have been established which are based upon the difference between radiated field measurements made at independent labs and Jerome Industries lab common mode current measurements on the cords of reference power supplies.
Using this technique, Jerome Industries has experienced a 100% passing rate for prototypes submitted for radiated field emissions testing at independent labs.

OEM System Engineer Instructions

*  The EMC characteristics of the final system comprised of both the power supply and the main system must be evaluated to the applicable EMC environment regulations.

*  Should a power supply need repair, please return it along with a note describing the particular problem to:

  Jerome Industries Inc.
  730 Division St.
  Elizabeth, NJ 07201
  U.S.A.

*  Circuit operation block diagrams, isolation diagrams, and a list of safety critical components will be available on request to appropriately qualified personnel.

* All products which contain double input fuses shall contain the following text in the system installation manual:
"CAUTION: DOUBLE POLE/NEUTRAL FUSING"

Medical Products

*  When used with medical products the Instruction Manual prepared for the end use product must incorporate at a minimum, the following power supply warning notices:
 

 a) Warning: When cleaning the power unit, use a cloth dampened with cleaning alcohol on the outside   of the enclosure only. Do not immerse product in water or a safety hazard could arise during use.
 b) Warning: Do not use this power unit in the presence of flammable anesthetics.
 

*  Periodic leakage current testing should be done on the combined power supply and end-use system on a yearly basis when used in a hospital environment where such test equipment is commonly available.


Appendix A


WSL400/WSL500 Family Typical Leakage Current
Measurements per IEC60601-1 Clause 19

  NC SFC
Earth Leakage Current 153.85 234.0
Enclosure Leakage Current 1.08 1.68
Patient Leakage Current
(to output negative cord connection)
19.38 86.69
Patient Leakage Current (floating),
Mains Voltage Applied (to output negative cord connection)
N/A 274.2

 

WSX400/WSZ400/WSZ500 Family Typical Leakage Current
Measurements per IEC60601-1 Clause 19

  NC SFC
Earth Leakage Current 234 420
Enclosure Leakage Current 0.9 1.7
Patient Leakage Current
(to output negative cord connection)
15 107
Patient Leakage Current (floating),
Mains Voltage Applied (to output negative cord connection)
N/A 383

 

Notes:
1) Measurement data in microamps
2) Measurements made with 275 Vac, 60 Hz input.
3) Worst case measurement condition, including humidity preconditioning.
4) Reduced leakage designs available (including CF category compliant designs), contact factory for availability.

 

Back Home