EXCEL FAMILY SWITCHMODE POWER SUPPLY SPECIFICATIONS

Specifications:

1. General Character:
The model WSL400, WSL500, and WSL900 series power supplies are designed for use in medical, information technology or laboratory  environments. They provide a "universal"  AC input operation.

1.1 The model number suffix "M", indicates that a specific model is for use in a medical environment.

1.2 Optional suffix indicates mechanical input configuration:

  N, North American (or NEMA) plug-in
  U, United Kingdom plug-in BS1363
  E, European mainland, CEE7/16
  A, Australian, AS3112
  H, Non-detachable Hospital Grade input cord
  If blank, then the standard IEC320 inlet is specified.

2. Electrical Specifications:

2.1 Input: 100-250Vac     50/60Hz
(90-264 Vac, 47-63 Hz is the tested operational range, 100-250 Vac range is marked on product to comply with safety standards.)

2.1.1 Input Fuses:
A current fuse is incorporated in each of the 2 AC lines, slo blo type fuse element, configured with integral welded on leads. Fuse Rating: WSL400 family = 1.25A, 250V / WSL500 family = 2A, 250V / WSL900 familiy = 1.25A, 250V.

2.1.2 Input Switch:
An optional input switch may be provided if required. A "I" and a "O" will be marked on it, or next to it.

2.2 Output:  
( Optional model number suffix letter "M", indicates medical usage. )

WSL433
WSL439
WSL445
WSL450
WSL455
WSL460
WSL465
WSL470
WSL475
WSL480
WSL490
WSL410
WSL411
WSL412
WSL413
WSL414
WSL415
WSL416
WSL418
WSL420
WSL422
WSL424
WSL426
WSL428
WSL430
WSL432
WSL436
WSL440
WSL444
WSL448
3.3Vdc 8.2A
3.9Vdc 8.2A
4.5Vdc 8.2A
5.0Vdc 8.2A
5.5Vdc 7.6A
6.0Vdc 7.1A
6.5Vdc 6.7A
7.0Vdc 6.2A
7.5Vdc 5.83A
8.0Vdc 5.5A
9.0Vdc 5.0A
10Vdc 4.7A
11Vdc 4.4A
12Vdc 4.2A
13Vdc 3.85A
14Vdc 3.55A
15Vdc 3.3A
16Vdc 3.1A
18Vdc 2.8A
20Vdc 2.5A
22Vdc 2.3A
24Vdc 2.1A
26Vdc 1.9A
28Vdc 1.8A
30Vdc 1.7A
32Vdc 1.6A
36Vdc 1.4A
40Vdc 1.25A
44Vdc 1.14A
48Vdc 1.04A
  WSL533
WSL539
WSL545
WSL550
WSL555
WSL560
WSL565
WSL570
WSL575
WSL580
WSL590
WSL510
WSL511
WSL512
WSL513
WSL514
WSL515
WSL516
WSL518
WSL520
WSL522
WSL524
WSL526
WSL528
WSL530
WSL532
WSL536
WSL540
WSL544
WSL548
3.3Vdc 10A
3.9Vdc 10A
4.5Vdc 10A
5.0Vdc 10A
5.5Vdc 9.4A
6.0Vdc 8.6A
6.5Vdc 8.0A
7.0Vdc 7.4A
7.5Vdc 6.9A
8.0Vdc 6.5A
9.0Vdc 6.0A
10Vdc 5.5A
11Vdc 5.0A
12Vdc 5.0A
13Vdc 4.6A
14Vdc 4.3A
15Vdc 4.0A
16Vdc 3.75A
18Vdc 3.3A
20Vdc 3.0A
22Vdc 2.7A
24Vdc 2.5A
26Vdc 2.3A
28Vdc 2.15A
30Vdc 2.0A
32Vdc 1.85A
36Vdc 1.65A
40Vdc 1.50A
44Vdc 1.35A
48Vdc 1.25A
  WSL933
WSL939
WSL945
WSL950
WSL955
WSL960
WSL965
WSL970
WSL975
WSL980
WSL990
WSL910
WSL911
WSL912
WSL913
WSL914
WSL915
WSL916
WSL918
WSL920
WSL922
WSL924
WSL926
WSL928
WSL930
WSL932
WSL936
WSL940
WSL944
WSL948
3.3Vdc 6.4A
3.9Vdc 6.4A
4.5Vdc 6.4A
5.0Vdc 6.4A
5.5Vdc 5.8A
6.0Vdc 5.3A
6.5Vdc 4.9A
7.0Vdc 4.6A
7.5Vdc 4.3A
8.0Vdc 4.0A
9.0Vdc 3.5A
10Vdc 3.3A
11Vdc 3.3A
12Vdc 3.3A
13Vdc 3.1A
14Vdc 2.8A
15Vdc 2.6A
16Vdc 2.5A
18Vdc 2.2A
20Vdc 2.0A
22Vdc 1.8A
24Vdc 1.6A
26Vdc 1.5A
28Vdc 1.4A
30Vdc 1.3A
32Vdc 1.25A
36Vdc 1.1A
40Vdc 1.0A
44Vdc 0.9A
48Vdc 0.8A

 

2.3 Output Variance, Voltage:
+/- 5% at Full Load, any input voltage or environmental condition

2.4 Load Regulation, Voltage:
Not to exceed 800 millivolts for 100% to 0% load change, varies due to output cord selected, remote sense option available, contact factory.

2.5 Maximum Output Ripple:
         3.3 - 6V: 50mVp-p
         6.1 - 24V: 100mVp-p
         24.1 - 36V: 150mVp-p
         36.1 - 48V: 200mVp-p

2.6 Protection Circuitry: The output is protected from short circuit and prolonged overload.

3. Mechanical Specifications:

3.1 Enclosure Dimensions:
4.75"L X 3.00"W X 1.90 "H inches. Enclosure outline drawings available upon request.

3.2 Input Configuration: Male IEC320 power inlet, 2 or 3 conductor option. Optional input connector lock available. Optional non-detachable input cords available for various international requirements.

3.3 Output Configuration:   6 foot length standard, various conductor arrangements and jacket material selections are available. Cordsets may be terminated with various dc connectors, custom output connector assemblies are available, contact customer service for details. Standard in-stock cordsets can be seen in the Jerome Industries engineering manual / catalog on pages 10 and 11.

4. Environmental:
 

4.1 Operating :

Temperature: 0 to 40C 
Humidity: 10% to 95% non-condensing
Atmospheric pressure: 70 to 106 kPa

Please contact customer service if a
wider operating range is required.

4.2 Storage :

Temperature: -40 to 75C
Humidity: 10% to 95% non-condensing
Atmosperic pressure: 50 to 106 kPa

5. Regulatory Compliance

5.1 Safety Compliance:
These power supplies have been certified by the following safety agencies to the standards shown:
 
5.1.1 Single Output Models
These power supplies have been certified by the following safety agencies to the standards shown:

Regulatory Agency

UL
 

 UL testing and 
evaluation of 
Canadian 
requirements.  

 

TUV
WSL400/500 Family


TUV
WSL900 Family

Standard

UL60601-1
UL60950-1 


C22.2#601-1

C22.2#60950-1



EN60601-1

EN60950-1

EN60601-1

EN60950-1

Classification

Isolation, BF
SELV


Isolation, BF

SELV



Isolation, BF

SELV

Isolation, BF

SELV

File#

E160908
E148586


E160908

E148586



B051124072021

B051124072021

B030624072011
B030624072011

5.2 Power supplies shipped to European market need to bear the CE mark. Jerome Industries can supply a CE mark under the following standards:
 
Standard

EN60601-1
EN60950-1
Classification

Isolation, BF
SELV

5.3 Electro Magnetic Compatibility:
Meets requirements of European EMC directive and FCC requirements.

5.3.1 Medical:
Meets EMC requirements from EN60601-1-2 (EMC requirements for ElectroMedical Equipment).

5.3.2 Emissions:
Meets level B requirements of FCC 47 CFR part 15 and CISPR11 (EN55011) and CISPR22 (EN55022), EN61000-6-3 (light industrial) and line harmonic emissions EN61000-3-1.

5.3.3 Immunity:
Meets CISPR24 (EN55024), EN61326 and EN61000-6-3.

5.3.4 EMC Responsibility:
It is the responsibility of the OEM customer to verify EMC compliance with his system before marketing the end-use system.
 

 Notes:

1) Emissions are monitored on all prototypes using a spectrum analyzer with peak, quasi peak and average mode power measurement capabilities. Conducted emissions are recorded for both 120 Vac and 230 Vac input conditions. A correlated radiated emissions program is used in order to economically verify the radiated emissions from a specific model.
The emissions program works from the premises of the power supply acting as a dipole antenna radiator, one side of the dipole antenna being the input cord, the other side being the output cord. Levels are monitored on the cords by a common mode current sensing device operating into a spectrum analyzer. Correlation factors have been established which are based upon the difference between radiated field measurements made at independent labs and Jerome Industries lab common mode current measurements on the cords of reference power supplies.
Using this technique, Jerome Industries has experienced a 100% passing rate for prototypes submitted for radiated field emissions testing at independent labs.

OEM System Engineer Instructions

*  The EMC characteristics of the final system comprised of both the power supply and the main system must be evaluated to the applicable EMC environment regulations.

*  Should a power supply need repair, please return it along with a note describing the particular problem to:

  Jerome Industries Inc.
  730 Division St.
  Elizabeth, NJ 07201
  U.S.A.

*  Circuit operation block diagrams, isolation diagrams, and a list of safety critical components will be available on request to appropriately qualified personnel.

* All products which contain double input fuses shall contain the following text in the system installation manual:
"CAUTION: DOUBLE POLE/NEUTRAL FUSING"

Medical Products

*  When used with medical products the Instruction Manual prepared for the end use product must incorporate at a minimum, the following power supply warning notices:
 

 a) Warning: When cleaning the power unit, use a cloth dampened with cleaning alcohol on the outside   of the enclosure only. Do not immerse product in water or a safety hazard could arise during use.
 b) Warning: Do not use this power unit in the presence of flammable anesthetics.
 

*  Periodic leakage current testing should be done on the combined power supply and end-use system on a yearly basis when used in a hospital environment where such test equipment is commonly available.


Appendix A


WSL400/WSL500 Family Typical Leakage Current
Class 1 Input Configuration

Measurements per IEC60601-1 Clause 19

  NC SFC
Earth Leakage Current 153.85 234.0
Enclosure Leakage Current 1.08 1.68
Patient Leakage Current
(to output negative cord connection)
19.38 86.69
Patient Leakage Current (floating),
Mains Voltage Applied (to output negative cord connection)
N/A 274.2

 

WSL900 Family Typical Leakage Current
Class 1 Input Configuration

Measurements per IEC60601-1 Clause 19

  NC SFC
Earth Leakage Current 161.55 243.8
Enclosure Leakage Current 0.89 1.57
Patient Leakage Current
(to output negative cord connection)
19.43 80.52
Patient Leakage Current (floating),
Mains Voltage Applied (to output negative cord connection)
N/A 240.8

 

Notes:
1) Measurement data in microamps
2) Measurements made with 275 Vac, 60 Hz input.
3) Worst case measurement condition, including humidity preconditioning.
4) Reduced leakage designs available (including CF category compliant designs), contact factory for availability.

 